SPECIFICATIONS

Metric

Order No. Range Remarks Mass
500-776 0 – 150mm with data output 180g
500-777 0 – 200mm 210g
500-774 0 – 150mm w/o data output 180g
500-775 0 – 200mm 210g

Inch/metric

Order No. Range Remarks Mass
500-786 0 – 6″ with data output 180g
500-787 0 – 8″ 210g
500-784 0 – 6″ w/o data output 180g
500-785 0 – 8″ 210g

DIMENSIONS

Range A B C D H L
0 – 150mm/0 – 6″ 16.5 21 13 40 16 233
0 – 200mm/0 – 8″ 20 24.5 16.5 50 16 290

Jaw thickness: 3.5mm

Optional accessories

(Dedicated for the models equipped with a digimatic output function (Code No. 500-776, 500-777, 500-786 and 500-787)) For details, refer to page D-39.

* Cannot be used for other than water resistant type Digital calipers with external output function.

About the charge function (SuperCaliper)

The minimum illumination required in the uncharged state is 60lux.
As shown in the table ‘JIS Z 9110 Artificial Illumination Intensity Standard’, this SuperCaliper can be used with confidence in a normal work environment.

The charge function allows the operator to use the SuperCaliper without interrupting work even if the ambient illumination is temporarily insufficient.

 

IP67 protection level
Level 6: Dust-proof.

No ingress of dust allowed.
Level 7: Protected against water penetration.
Ingress of water in quantities causing harmful effects shall not be possible when the enclosure is temporarily immersed to a depth of 1 meter in water under standardized conditions of pressure and time (30 min.).

Demonstration device

Technical Data
Resolution: 0.01mm or .0005″/0.01mm Accuracy: ±0.02mm (excluding quantizing error) Repeatability: 0.01mm

Quantizing error: ±1 count Dust/Water protection level: IP67* Power supply: Solar cell**
Display: LCD
Scale type: ABSOLUTE electromagnetic induction linear encoder
Max. response speed: Unlimited
* This model is not waterproof type. Therefore, rustproofing shall be applied after use.
** Can be used continuously above 60lux ambient illumination.

Functions

Origin-set: Absolute origin position can be changed.
Alarm: Error message is displayed and measurement

functions become inoperative if:

Air leak test equipment for water resistance inspection
Generally, an air leak test is adopted for evaluating water resistance.

Procedure: Place the measuring tool inside the capsule and seal it. Then fill the capsule and the test chamber with air at the required pressure and close the isolating valves. If there is no leak in the measuring tool, the differential pressure sensor will read zero, because the amount of air inside the test chamber is unchanging. However, if there is a leak in the measuring tool, the differential pressure sensor will show a non-zero reading due to a decrease in pressure inside the test chamber as air leaks into the tool. By detecting this differential pressure, GO/NG judgment for the severity of the leak is performed. This air leak test is performed for all ABS coolant proof calipers and coolant proof micrometers.

Air leak test equipment for ABS coolant proof caliper